maIntenance
109
The
Tip clean
and
Tip rinse
solutions are needed to clean and rinse the probes
after measurement, before the system stops and should be replaced (if not used
up) all three months.
The
Cuvette
Clean
is used during run for tip pre/post wash- and cuvette post
wash and should be replaced (if not used up) all three months.
The
Diluent
is used for sample/calibrator pre- and post dilution and for blank
measurements and should be replaced (if not used up) once a month.
1. Check the system wash solution behind the front door, if necessary prepare
new, for details refer to the IFU of Wash Add Ref 18971. After replacement
flush the system. Don’t top up to avoid contamination and bacterial grow.
2. Check the Tip Clean and - Rinse (Ref 16663/25) in the reagent tray, replace
if necessary.
3. Check the Cuvette Clean (Ref. 16663/20) in the reagent tray, replace if nec-
essary.
4. Check the Diluent (Ref. 16663/10) in the reagent tray, replace if necessary.
8.2.3 EMPTY WASTE BOTTLE
The waste bottle is placed on a scale and if the bottle is full, the system will show
an alarm. Open the front door, empty the waste bottle and put it back into the
system.
8.2.4 HYDRAULIC TESTING – SYSTEM FLUSH
System flush is required at least once per day and can be set to be automatically
performed during analyzer start up at the beginning of the day. In case that
analyzer has stayed operational and has not been switched off then the system
flush warning will come up at the specified time. Air gaps and bubbles should
be flushed, if present, particularly after the filling operation of system bottles or
after cleaning the probes.
1. Open the Maintenance scheduler status screen and press
Perform
behind
the
System Flush
function.
2. During process look for:
- Presence of bubbles or air gaps in system
- Leakage in peristaltic pump
- Constant and uniform flow from probe tip
- No droplets hanging on probe tip
- In case new bubbles generate in the process, Refer to chapter 9 “Trouble-
shooting”.
3. The scheduler will be updated automatically.
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......