HumaStar 600
| User manual
46
5.1.5 REFILLING REAGENT BOTTLES (ONLY FOR OPEN CHANNELS)
When pressing right button over a given reagent, menu will include a Refill
option. Once the
Apply Changes
button is pressed, tray will move in the desired
reagent position.
Several reagent vials can be refilled with this procedure: when
Apply Changes
is pressed tray will sequentially position in all the defined refilling reagents. The
dilution restart is automatic.
5.1.6 METHOD ASSIGNMENT TO TRAYS
Methods can be assigned for processing either in the front or back reaction tray.
Only those methods defined as Methods in Use can be assigned to a tray. De-
fault assignment is to the front tray if not otherweise specified at the applica-
tion protocol of the method.
For tray assignment, follow:
There is a unique feature consisting an automatic reassignment of methods to
the trays, according historic statistics and with the criterion of minimum total
analysis time. Automatic reassignment does not include interfering or inter-
fered methods. Interfered methods are posted in red color and consecutive use
is avoided either at the arm or cuvette sequence level. To apply it, first
Refresh
the desired period for statistical analysis and then press the
Automatic
button.
Moved methods will be indicated with an X. Press
Apply
to save the changes.
5.2 Samples
Samples can be loaded directly or associated to patients. Tests and calculated
methods (which including a chemistry measurement) are assigned to samples
Before starting automatic
procedure, volumes can be
tested. Press Check Volumes
button for this operation.
FIGURE 47
To remove a diluent or cleaning solution from the tray press
Remove Solution
,
select or type in the desired solution and proceed as above.
To select more than one item from the list, press and hold
Ctrl
key while select-
ing the new item. To select a range of items, select the first item, then press and
hold
Shift
key while selecting the last item. Alternatively for extended selection
use the mouse by click and drag.
Alternatively, to remove a reagent or solution from a particular position in
the tray, right click the desired position and pick
Remove
option.
Press
Apply Change
s
and the Reagent tray drive the selected position/s to
the bottle insert/remove area. Following window appears:
Open the cover, take out the reagent or solution bottles and close the cover.
Press
Ok
button to confirm the operation.
Same procedure must be followed for removing diluents and solutions.
FIGURE 45
FIGURE 46
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......