HumaStar 600
| User manual
124
Reagents
Maximum number of simultaneous tests: 24 double to
48 single reagent tests + 3 with optional ISE unit.
1 to 3 reagents, 5 to 500 µL/test each (in increments of
1 µL), final total solution volume 180 to 500 µL/test
Reagent bottles capacities: 25 and 70.
Reagent cooling compartment: 48 cooled positions.
Multiple vials per test.
Reagent temperature 9°C+/-2°C for room temperature
between 15 and 30°C.
Reaction
Water consumption: 5 L / hour.
Warm air incubator: 37°C.
Reaction cuvette: re-usable plastic 6 mm light path
with 6-stage washing
Reaction time: 0 to 10 min.
Reaction temperature: 37°C ± 0.1°C.
Stirring: After dispensing each reagent.
Optics
Double beam
Photometric Range: -0.1 to 3.6 A.
Measuring wavelength: 340 to 800 nm (selectable
among 12 wavelengths). Photometry: Single or Dou-
ble-wavelength simultaneous reading.
ISE Unit
Na
+
, K
+
and Cl
-
measurements.
Samples: serum or urine.
Data Management
Windows
TM
based Software.
Interface LIS: bi-directional RC 232 C, according to ASTM
1394 requirements.
Printout
Customer’s optimized (Analysis result, work list, list of
samples, Quality
Control, Calibration curves, etc.)
Environmental conditions
Transportation and storage: 10 to 40°C, humidity 30 to
90%, pressure 600 to 1050 hPa.
Use: 15 to 30 °C, humidity 40 to 80%, pressure 600 to
1050 hPa.
Power Requirements
110/220V, 50/60 Hz, 1,4 kVA
Dimensions (WxDxH):
Instrument without
any components:
100 x 74 x 113 cm
Space required for
routine use:
162 x 128 x 160 cm
Packaging:
135 x 86 x 135 cm
Weight:
Gross: 250 kg, Net: 180 kg.
Summary of Contents for HumaStar 600
Page 1: ...HumaStar 600 User Manual Cat No 16660 001...
Page 2: ......
Page 4: ......
Page 14: ...HumaStar 600 User manual 10...
Page 44: ...HumaStar 600 User manual 40...
Page 108: ...HumaStar 600 User manual 104...
Page 129: ...125...
Page 130: ...HumaStar 600 User manual 126...
Page 131: ......