2.12 None Of The Electrical Functions Operate
Chapter 2: Troubleshooting Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 2 - 45
2
4. The siderail control cables (A) (if available) and the handset control cable
(B) are firmly connected to their respective receptacles on the control
lockout box and are in good condition (see figure 2-32 on page 2-42).
Yes
No
↓
→
Connect the cables, and replace any that are damaged.
This solves the problem.
Yes
No
↓
→
Call Hill-Rom
®
Technical Support at (800) 445-3720.
5. Press the Up and Down buttons for each function on the handset and on the
optional siderail controls (if available). This creates a relay clicking sound
in the control box (on top of the hilow motor actuator).
Yes
No
↓
→
If a clicking sound is heard with one control actuated but not the
other(s), go to RAP 2.13. If a relay clicking sound is not heard
when any of the controls are pressed, replace the control box
(see “Control Box” on page 4-35).