![Hill-Rom P870 Manual Download Page 140](http://html1.mh-extra.com/html/hill-rom/p870/p870_manual_2134201140.webp)
4.10 Control Lockout Box
Chapter 4: Removal, Replacement, and Adjustment Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 4 - 33
4
4.10 Control Lockout Box
Tools required:
Phillips head screwdriver
Diagonal cutters
Removal
1. Unplug the bed from its power source.
SHOCK HAZARD:
Unplug the bed from its power source. An electrical shock hazard
exists.
2. Disconnect, if applicable, the left-hand siderail cable (A), right-hand
siderail cable (E), handset cable (D), and the automatic contour cable (C)
from the control lockout box (see figure 4-24 on page 4-33).
Figure 4-24. Location Of Control Lockout Box Cable Connections
3. Disconnect the power/hilow cable (B) from the control box mounted on top
of the hilow motor actuator (see figure 4-14 on page 4-17).
NOTE:
Record the cable routing and placement of cable ties for replacement purposes.
4. Remove the attached cable ties.
5. Remove the three screws (A) mounting the control lockout box to the foot
of the main frame (see figure 4-25 on page 4-34). Remove the control
lockout with the attached cable as an assembly.
A
B
C
D
E
m136_071
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...