2.10 Automatic Contour Does Not Function Correctly
Chapter 2: Troubleshooting Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 2 - 39
2
4. Pressing the HEAD Down button with the head section inclined above 15
o
causes only the head section to lower. Continuing to press the HEAD
Down button causes the knee section to begin lowering. Lowering of both
sections continues (with HEAD Down pressed) until the flat position is
reached. Releasing the HEAD Down button before the flat position is
reached causes both sections to stop moving.
Yes
No
↓
→
Go to step 5.
5. The automatic contour cable (C) is firmly connected to its receptacle on the
control lockout box. Routing to the automatic contour module shows no
visible damage.
Yes
No
↓
→
Plug the cable firmly into its receptacle, or replace the damaged
cable if necessary.
6. Press inward on the knee plunger (A) of the automatic contour module with
the HEAD Up button depressed (see figure 2-30 on page 2-39). The knee
section rises.
Yes
No
↓
→
Replace or repair the automatic contour module (see “Automatic
Contour Module” on page 4-22).
Figure 2-30. Automatic Contour Module
A
m136_029
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...