6.1 Leakage Current Test for Electric Drive Model
Chapter 6: General Procedures
Page 6 - 13
Resident
TM
LTC Bed Service Manual (man136rc)
6
Figure 6-1. Double Insulated Product Test Adapter
6. Turn on the leakage analyzer, and set the function knob to A case
leakage.
7. Toggle the ground and polarity switches to obtain eight separate readings.
8. Record the current leakage that is displayed on the analyzer in table 6-5 for
each setting.
Table 6-5. Current Leakage Test Values
Unit On:
G/NP:
NG/NP:
G/RP:
NG/RP:
Unit Off:
G/NP:
NG/NP:
G/RP:
NG/RP:
m136_100
µ