
4.6 Knee Section Motor Actuator
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 20
Resident
TM
LTC Bed Service Manual (man136rc)
4
4.6 Knee Section Motor Actuator
Tools required: E-ring pliers Pin punch
Hammer Pliers
Diagonal cutters Lithium grease (P/N SA3351)
Removal
1. Remove the mattress, and completely lower the head section of the bed.
2. Unplug the bed from its power source.
SHOCK HAZARD:
Unplug the bed from its power source. An electrical shock hazard
exists.
3. Manually raise the head section of the bed, and unplug the knee motor
actuator cable (A) from the control box (see figure 4-16 on page 4-20).
Figure 4-16. Knee Section Motor Actuator
A
B
B
C
m136_062
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...