TM
LTC Bed Service Manual (man136rc)
4
Figure 4-25. Control Lockout Box Mounting Screws
Replacement
1. Assemble in reverse order.
2. Ensure that the power/hilow cable is properly routed and connected to the
control box (mounted on top of the hilow motor actuator).
3. Install the cable ties to secure the power/hilow cable where it will not be
damaged during bed operation.
4. Connect the siderail cables, handset cable, and automatic contour cable to
their respective receptacles on the control lockout box.
5. Perform an operational check of the hilow, head, and knee functions. See
“Function Checks—Electric Drive Model” on page 2-10.
m136_072
A