2.8 Electric Head Section Does Not Raise Or Lower Correctly
Chapter 2: Troubleshooting Procedures
Page 2 - 34
Resident
TM
LTC Bed Service Manual (man136rc)
2
5. The head section rises when the HEAD Up button is pressed and stops
when the button is released. Press and hold the HEAD Up button. This
adjusts the incline of the head section up to its limit of approximately 65
o
.
Yes
No
↓
→
6. Press the HEAD Down button. This adjusts the incline of the head section,
and travel stops when the button is released. Press and hold the HEAD
Down button. This adjusts the head section down to the flat position.
Yes
No
↓
→
7. Turn the control lockout knob for the head function to the locked position.
This prevents head operation. Turn the control lockout knob for the head
function to the unlocked position. This allows the head section to be
operated up and down.
Yes
No
↓
→
8. Go to “Final Actions—Electric Drive Model” on page 2-18.
9. Another part of the bed interferes with the head section movement to the
maximum incline or flat position.
Yes
No
↓
→
Replace the head motor actuator (see “Head Section Motor
Actuator” on page 4-18).
10. Eliminate the interference with the head movement of the bed.
This solves the problem.
Yes
No
↓
→
Call Hill-Rom
®
Technical Support at (800) 445-3720.
11. Go to “Final Actions—Electric Drive Model” on page 2-18.
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...