4.1 Manual Hilow Drive Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 4 - 5
4
4.1
Manual Hilow Drive Assembly
Tools required:
Crescent wrench
T25 torx head screwdriver
Needle nose pliers
Jack stands
Socket wrench
5/8" deep-well socket
Phillips head screwdriver
Light oil (P/N 8252)
Removal
WARNING:
Only facility-authorized maintenance personnel should perform
maintenance procedures on the Resident
TM
LTC bed. Otherwise,
personal injury can occur.
1. Remove the mattress from the bed.
2. Set both brakes on the bed by fully depressing the brake pedals (A) located
on both sides of the swivel caster frame assembly at the foot end of the bed
(see figure 4-1 on page 4-5). If the brake design has interconnecting
linkage, both brakes are applied when either pedal is depressed. Give the
bed a solid tug to ensure the brakes are set.
Figure 4-1. Brake Pedal Assembly
3. Adjust the bed sleep surface to its lowest position by turning the hilow
crank handle (middle handle) counterclockwise until it stops (see figure 4-
2 on page 4-6).
WARNING:
Electric or manual drive bed mechanisms can cause serious injury if
operated improperly. Operate the bed only when persons are clear of
A
m136_010
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...