4.11 Control Box
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 36
Resident
TM
LTC Bed Service Manual (man136rc)
4
4. Slide the plastic door (D) towards the center of the bed, and remove this
door (for access to a hidden mounting screw).
5. Remove the hidden mounting screw (securing the control box to the hilow
motor actuator.
6. Slide the control box toward the center of the bed, and remove it from the
hilow motor actuator.
Replacement
1. Assemble in reverse order.
2. Ensure that the base of the control box correctly engages its mounting
slides on top of the hilow motor actuator. The hilow motor actuator
connecting pin must align with its receptacle in the control box.
CAUTION:
The hilow motor actuator connecting pin must align with its receptacle
in the control box. Equipment damage may occur.
3. Ensure that all cables are properly connected to the control box.
4. Perform an operational check of the hilow, head, and knee functions. See
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...