8. Unplug the bed from its power source or unplug the handset control cable
(B) from the control lockout (if the optional siderail controls are installed,
unplug one of the siderail controls (A) from the control lockout).
9. Identify the location of all wire ties with a white marker, and cut the wire
ties holding the control cables to the metal channel. Unplug the control
lockout cable (A) from the control box, and plug the disconnected handset
(or siderail control) in its place (see figure 2-33 on page 2-43).
10. Plug bed into an appropriate power source, and press control buttons to
operate the bed’s functions. All motor actuators operate in both directions,
but not as indicated on the control (i.e.: head may be operated by the HI/LO
buttons, etc.).
SHOCK HAZARD:
Unplug the bed from its power source before unplugging any cables
from the control lockout, or from the control box. An electrical shock
hazard exists.
Yes
No
↓
→
Call Hill-Rom
®
Technical Support at (800) 445-3720.
Figure 2-33. Control Lockout Cable Connection at Control Box
11. Replace control lockout box (see “Control Lockout Box” on page 4-33).
12. Go to “Final Actions—Electric Drive Model” on page 2-18.
A
m136_032
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...