4.5 Head Section Motor Actuator
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 18
Resident
TM
LTC Bed Service Manual (man136rc)
4
4.5
Head Section Motor Actuator
Tools required:
E-ring pliers
Pin punch
Hammer
Pliers
Diagonal cutters
Lithium grease (P/N SA3351)
Removal
1. Remove the mattress, and completely lower the head section of the bed.
2. Unplug the bed from its power source.
SHOCK HAZARD:
Unplug the bed from its power source. An electrical shock hazard
exists.
3. Manually raise the head section of the bed, and unplug the head motor
actuator cable (A) from the control box (see figure 4-15 on page 4-18).
Figure 4-15. Head Section Motor Actuator
A
B
C
B
m136c060