6.1 Leakage Current Test for Electric Drive Model
Chapter 6: General Procedures
Page 6 - 12
Resident
TM
LTC Bed Service Manual (man136rc)
6
6.1
Leakage Current Test for Electric Drive Model
NOTE:
The following leakage current test is used only for those units that do not have
the factory installed ground wire. For those units that have the factory installed
wire, perform a standard leakage current test.
The leakage current test is designed to test the level of patient environment
safety. Maximum allowable electrical current for the Resident
TM
LTC electric
drive bed is 300 A per UL 2601-1. Perform the following test using a double
insulated product test adapter (see figure 6-1 on page 6-13).
NOTE:
Consult the test equipment manufacturer to obtain a product test adapter.
1. Unplug the bed from its power source.
2. Plug the leakage test analyzer into the AC wall outlet.
3. Plug the male (A) connector of the double insulated product test adapter
into the test receptacle of the analyzer.
4. Plug the Resident
TM
LTC electric drive bed power cord into the female
connector (B) of the double insulated product test adapter.
SHOCK HAZARD:
One of the risks associated with the use of electrical equipment
includes the potential for electrical shock. Train and educate personnel
on the risks associated with electrical equipment.
5. Attach the alligator clip lead (C) to an exposed non-painted screw that is
screwed into the bed frame at the head end.
µ
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...