![Hill-Rom P870 Manual Download Page 123](http://html1.mh-extra.com/html/hill-rom/p870/p870_manual_2134201123.webp)
4.4 Hilow Motor Actuator
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 16
Resident
TM
LTC Bed Service Manual (man136rc)
4
4.4
Hilow Motor Actuator
Tools required:
E-ring pliers
Pin punch
Hammer
Pliers
Diagonal cutters
Lithium grease (P/N SA3351)
Removal
1. Remove the mattress, and manually raise the head section of the bed to
gain access to the hilow motor actuator.
2. Use jack stands to block up beneath the bed to maintain its hilow position
when removing the hilow motor actuator from the bed.
3. Unplug the bed from its power source.
SHOCK HAZARD:
Unplug the bed from its power source. An electrical shock hazard
exists.
NOTE:
Record the cable routing and placement of cable ties for replacement purposes.
4. Disconnect all cables from the control box assembly (A) on top of the
hilow motor actuator (see figure 4-14 on page 4-17).