4.2 Manual Head Drive Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 4 - 9
4
4.2
Manual Head Drive Assembly
Tools required:
Needle nose pliers
Phillips head screwdriver
Crescent wrench
Lithium grease (P/N SA3351)
Removal
WARNING:
Only facility-authorized maintenance personnel should perform
maintenance procedures on the Resident
TM
LTC bed. Otherwise,
personal injury can occur.
1. Remove the mattress from the bed.
2. Set both brakes on the bed by fully depressing the brake pedals located on
both sides of the swivel caster frame assembly at the foot end of the bed. If
the brake design has interconnecting linkage, both brakes are applied when
either pedal is depressed. Give the bed a solid tug to ensure the brakes are
set.
3. Adjust the bed sleep surface to its highest position by turning the hilow
crank handle (middle handle) clockwise until it stops (see figure 4-7 on
page 4-9).
Figure 4-7. Manual Bed in the High Position
m136_108
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...