2.8 Electric Head Section Does Not Raise Or Lower Correctly
Chapter 2: Troubleshooting Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 2 - 33
2
Figure 2-24. Head Controls
4. The head motor actuator cable (A) is firmly connected to its receptacle on
the control box (on top of the hilow motor actuator), and the routing to the
control lockout box shows no visible damage (see figure 2-25 on page
2-33).
Yes
No
↓
→
Plug the cable firmly into its receptacle, or replace the damaged
cable if necessary.
This solves the problem.
Yes
No
↓
→
Replace the head motor actuator (see “Head Section Motor
Actuator” on page 4-18).
Figure 2-25. Head Cable Routing Between the Control Box and the Control Lockout
A
B
B
A
m136_023
A
m136_024