2.7 Electric Hilow Does Not Raise or Lower Correctly
Chapter 2: Troubleshooting Procedures
Page 2 - 30
Resident
TM
LTC Bed Service Manual (man136rc)
2
Figure 2-21. Hilow Controls
4. The power/hilow cable (A) is firmly connected to its receptacle on the
control box, and the routing to the control lockout box shows no visible
damage (see figure 2-22 on page 2-30).
Yes
No
↓
→
Plug the cable firmly into its receptacle, or replace the damaged
cable if necessary.
This solves the problem.
Yes
No
↓
→
Replace the hilow motor actuator assembly (see “Hilow Motor
Actuator” on page 4-16).
Figure 2-22. Hilow Cable Routing Between the Control Box and the Control Lockout
A
B
A
B
m136_019
m136_020
A
m136_021