![Hill-Rom P870 Manual Download Page 206](http://html1.mh-extra.com/html/hill-rom/p870/p870_manual_2134201206.webp)
Preventive Maintenance
Chapter 6: General Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 6 - 7
6
Preventive Maintenance Schedule—Manual Drive Model
Table 6-1. Preventive Maintenance Schedule—Manual Drive Model
Function
Procedure
Hilow limits
Crank the hilow function to the upper and lower limits to ensure
proper function of the hilow drive assembly. Lubricate the drive
mechanism.
Head limits
Crank the head function from the flat to its maximum angle of
60
o
(± 2
o
) to ensure proper function of the head drive assembly.
Lubricate the drive mechanism.
Knee limits
Crank the knee function from the flat to its maximum angle of
45
o
(± 2
o
) to ensure proper function of the knee drive assembly.
Lubricate the drive mechanism.
Brakes
Test the brakes to determine if the bed moves when the brakes
are activated.
Casters
Test the swivel casters to ensure they do not bind. Check all
caster tires for cuts, wear, tread, etc. Replace if necessary.
Siderails
Inspect for proper up, down, and storage operation.
Inspect the locking latch for proper operation.
Lubricate the mechanism.
Pivot points
Lubricate all pivot points on the bed.
Head and foot
panels
Check aesthetics. Clean and touch up if necessary.
Overall appearance Inspect the condition of the labels, paint, and general aesthetics.
Replace labels, touch-up paint, and clean if necessary.
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 201: ...Chapter 6 General Procedures Page 6 2 ResidentTM LTC Bed Service Manual man136rc 6 NOTES...