2.8 Electric Head Section Does Not Raise Or Lower Correctly
Chapter 2: Troubleshooting Procedures
Page 2 - 32
Resident
TM
LTC Bed Service Manual (man136rc)
2
2.8
Electric Head Section Does Not Raise Or Lower Correctly
1. The bed is plugged into the appropriate power source.
Yes
No
↓
→
Connect a good power cord to the wall outlet (see figure 2-34 on
page 2-44).
2. Ensure that the control lockout knob for the head function (A) is in the
unlocked position (see figure 2-23 on page 2-32).
Yes
No
↓
→
Turn the knob to the unlocked position.
Figure 2-23. Control Lockout Box Head Function
3. Press the HEAD Up (A) or the HEAD Down (B) button on the handset or
optional siderail controls (if available) (see figure 2-24 on page 2-33). This
creates a relay clicking sound in the control box on top of the hilow motor
actuator.
Yes
No
↓
→
If a clicking sound is heard with one control actuated but not the
other(s), go to RAP 2.13. If a relay clicking sound is not heard
when any of the controls are pressed, go to RAP 2.11 step 8.
A
m136_022