2.12 None Of The Electrical Functions Operate
Chapter 2: Troubleshooting Procedures
Page 2 - 44
Resident
TM
LTC Bed Service Manual (man136rc)
2
2.12 None Of The Electrical Functions Operate
1. The bed is plugged into the appropriate power source.
Yes
No
↓
→
Connect the power cord (A) to the wall outlet (see figure 2-34 on
page 2-44).
Figure 2-34. Power Cord Location At the Control Box Assembly
2. All cables are firmly connected to the control box assembly.
Yes
No
↓
→
Connect the cables to their respective receptacles.
3. Ensure that the control lockout knob for each of the operating functions is
in the unlocked position (see figure 2-31 on page 2-41).
Yes
No
↓
→
Turn the knobs to the unlocked position.
A
m136_033
Summary of Contents for P870