MAN ECG preamplifier
22.
MAN ECG preamplifier
22.1 Description
•
The ECG module consists of a ECG preamplifier type MAN 6 (hardware part) and a patient
connection cable (KENDALL medizinische Erzeugnisse, Code 8/P93/07-01, Ref 3227.0701.00).
•
The connector for the patient connection cable is on the front cover of the hardware section, which
is placed in a drive-slot in the front of the ultrasound machine.
•
The ECG preamplifier type MAN6 is used for acquiring an ECG signal to be displayed with the
ultrasound image. The ECG preamplifier must not be used for ECG diagnostics. It is not intended
for use as a cardiac monitor.
•
The signal input of the ECG preamplifier is equipped with a protection against high voltages used
for defibrillation (Type BF).
•
The ECG preamplifier is connected to a connector on the rear panel of Voluson
®
730Pro.
review also: Technical Data/Information:
MAN ECG preamplifier
(chapter 23.19)
Voluson
®
730Pro - Operation Manual
22-2
105831 Rev. 0
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...