Technical Data / Information
Ambient temperature: 10°C to 40°C resp. 50°F to 104°F (operation temp. of instrument)
-10°C to 40°C resp. 14°F to 104°F (stock and transport temp.)
working temp. of probes see probe user’s manual
Barometric pressure: 700 to 1060 hPa (operation condition)
700 to 1060 hPa (stock and transport condition)
Humidity:
30 to. 80% RH no condensation (operation condition)
0 to. 90% RH no condensation (stock and transport condition)
Humidity protection: covered, no humidity protection
Over-temperature protection: yes, provided by three fans and alert display
Dimensions:
680 x 1000 x 1427 mm [width x depth x height]
26.7 x 39.4 x 56.1 in.
Weight:
basic unit (without accessories) approx. 134 kg
23.2 Transmitter
Frequency range:
Broad band system 1 to 15 MHz, automated adaptation to the probe
used.
Control of acoustic output:
Range: 20dB, adjustable in 1dB steps
Focusing:
selectable transmit focal length and focal depth
Processing-channels:
256 channels
Sound field parameters:
the declaration of the sound field parameters acc. to IEC1157 can be
furnished by the manufacturer on request. Sound field limiting
parameters on request.
23.3 Receiver
Frequency range:
Broad band system 1 to 15 MHz, automated adaptation to the probe used.
Focusing (with annular and multi-element):
Sub-pixel based digital dynamic focusing system:
Accuracy of focus: +/- 3ns
Sample-rate:
60.0
MHz
Processing-channels:
High resolution mode: 256 channels
Receive-apodization:
yes
TGC:
manual, 100 dB control range by gain knob and slider-pots
Dynamic range:
150dB
Voluson
®
730Pro - Operation Manual
23-4
105831 Rev. 0
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...