Volume Mode
Start situation
Rotation (about X-axis)
The position of the volume body in relation to the display plane is determined by a relative coordinate
system. This is made up of three orthogonal axes, the common intersection of which is the 3-axial
center of rotation. These axes are displayed within the display plane - exactly in the
X-, Y- and Z-directions and colored. Rotation about any of these axes and displacement of the center
of rotation make any imaginable plane within the volume body displayable. The INIT position of the
volume body in relation to the display plane is resetable; it is the start situation after completion of a
volume scan.
The standard representation: 3 sections mode
The 3 orthogonal sectional planes are simultaneously displayed on the quartered screen field of the
monitor. Each quarter field displays a sectional view through the volume body as shown below.
Sectional planes A, B, C
Display of A, B, C
The intersection lines of the planes are displayed in colors:
AB = blue
AC = red
BC = yellow
Orientation of intersection lines on the screen:
Section/field
A
B
C
Intersection
line
AB
V V N
V
=
Vertical
Intersection line AC
H
N
H
H = Horizontal
Intersection line BC
N
H
V
N = Normal
By this definition the relation of the position of the 3 images A, B, C is also defined (as made clear by
the direction of arrows).
The presentation of 3 orthogonal sectional planes may lead to non-conformance with the conventional
customized orientation to the patient in 2D sonography. An identification system - the automatic
display of the direction of section - will clarify.
Please note:
Whenever a usual longitudinal section (of the patient) is selected for display field A,
the usual orientation for longitudinal and transverse sections is valid.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
11-17
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...