Connections
21.3 Important
Notes:
Connecting Auxiliary Equipment
CAUTION !
The leakage current of the entire system including any / all auxiliary
equipment must not exceed the limit values as per EN60 601-1-1:1990 (IEC
60601-1-1) resp. other valid national or international standards. All
equipment must comply with UL, CSA and IEC requirements.
CAUTION !
Auxiliary equipment must only be connected to the main console with the
special mains outlet provided for the electrical safety of the system.
CAUTION !
Auxiliary equipment with direct mains connection requires galvanic
separation of the signal and / or control leads.
CAUTION:
The outlets ST1 – ST5 may be set to 230V or 115V
Modification of voltage setting only by an authorized service person!
ATTENTION ! With the color video monitor connected to its isolated mains outlet, the
remaining load capacity for auxiliary equipment is max. 350 VA.
WARNING !
Only accessories explicitly recognized by the system manufacturer
GE Medical Systems Kretztechnik GmbH & Co OHG may be used in
connection with the system.
Voluson
®
730Pro - Basic User Manual
21-18
105831 Rev. 0
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...