Probes and Biopsy
20.12.4 Coupling
Gels
Do not use unrecommended gels (lubricants).
They may damage the probe and void the warranty.
Applying:
In order to assure optimal transmission of energy between the patient and probe, a conductive gel or
couplant must be applied liberally to the patient where scanning will be performed.
Precautions:
Coupling gels should not contain the following ingredients as they are known to cause probe damage:
•
Methanol, ethanol, isopropanol, or any other alcohol-based product
•
Mineral oil
•
Iodine
•
Lotions
•
Lanolin
•
Aloe Vera
•
Olive Oil
•
Methyl or Ethyl Parabens (para hydroxybenzoic acid)
•
Dimethylsilicone
20.12.5 Planned
Maintenance
The following maintenance schedule is suggested for the system and probes to ensure optimum
operation and safety.
Do the Following
Daily
After Each Use
As Necessary
Inspect the Probes
X
Clean the Probes
X
Disinfect Probes
X
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
20-19
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...