Probes and Biopsy
20.8 Probe
Usage
For details on connecting, activating, deactivating, disconnecting, transporting and storing the probes,
refer to:
Transducer Connection
(chapter 4.4)
and
Probe/Program Selection
(chapter 4.5).
20.9 Care and Maintenance
20.9.1 Inspecting
Probes
After each use, inspect the probe's lens, cable, and casing. Look for any damage that would allow
liquid to enter the probe. If any damage is found, do not use the probe until it has been inspected and
repaired/replaced by a GE Medical Systems - Kretztechnik Service Representative.
NOTE:
Keep a log of all probe maintenance, along with a picture of any probe malfunction.
20.9.2 Environmental
Requirements
Probes should be operated, stored, or transported within the parameters outlined below.
Ensure that the probe face temperature does not exceed the normal
operation temperature range.
Probe Environmental Requirements
Operational Storage Transport
Temperature
+10° to +40°
C
+50° to +104° F
-10° to
+60°
C
+14° to +140° F
-40° to
60°
C
-40° to +140° F
Humidity
Max. 90%
non-condensing
Max. 90%
non-condensing
Max 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Voluson
®
730Pro - Basic User Manual
20-8
105831 Rev. 0
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...