Calculations and Patient Reports
14.7.4.4 LV
Mass
This is used for the measurement of the Left Ventricular volume and mass.
It is correctly measured only during the diastolic phase (LV expanded).
1. Press
the
[
Calc
]
key.
The Cardiac calculations menu appears in the menu area.
2. Select
[
LV Mass
]
3. Perform the measurement of
[
Epicardial Area
]
,
[
Endocardial
Area
]
and
[
Epicardial Length
]
prior to the LV mass measurement.
4. When the cursor appears in the image, measure the selected item
using the trackball and the right or left trackball key
[
Set
]
.
5. To obtain the second, orthogonal image, press
[
Freeze
]
.
Scan the image and press
[
Freeze
]
again.
Note:
When using Dual 2D mode, it is not necessary to unfreeze during measurements.
14.7.4.5
LVOT- or RVOT Diameter
Note:
This 2D measurement is part of the LVOT / RVOT Doppler measurements.
To calculate LVOT- or RVOT diameter in 2D mode:
1. Press
the
[
Calc
]
key.
The cardiac calculations menu appears in the menu area.
2. Select
[
LVOT
]
or
[
RVOT
]
and
[
LVOT Dia
]
or
[
RVOT Dia
]
.
3. Perform the measurement by using the trackball and the right or
left trackball key
[
Set
]
.
Voluson
®
730Pro - Basic User Manual
14-30
105831 Rev. 0
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...