Volume Mode
Rotation about the Z-axis of the reference image (e.g.
A
).
Clockwise turn of the
[
Z-axis
]
rotary control :
By the clockwise rotation of the volume body in relation to the screen plane (as shown) the new
sectional planes are calculated in real time and displayed on screen.
Important notes for the user:
•
Rotation should be performed slowly to understand the orientation.
•
Do not rotate by large angles except when the orientation left/right or up/down is to be changed.
At 90° rotation about an axis, the sections A, B, C will change:
•
Reference image e.g., A: X-axis: A
↔
C
Y-axis: A
↔
B
Z-axis: B
↔
C
•
Before performing a rotation, position the center of rotation at that part of the image which you
don't want to lose.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
11-21