TD Mode
10.3.8 Balance
“Balance” controls the amount of Color displayed over bright echoes and helps confine Color within
the vessel walls. Decreasing this Balance displays Color on brighter structures.
If you see no color, the Balance is probably set too high.
Rotate the
[
Balance
]
digipot and select the useful Balance.
The Balance orientation line is only visible in the color modes.
The line represents to the position of the adjusted displayed gray
value in the gray scale.
Remarks:
*
Changes are only possible in write mMode !
10.3.9 Dynamic
Set
“Dynamic Set” controls the axial resolution of color in the display.
It adjusts the axial sample depth of color pixels.
When rotating counter-clockwise the color information is sharper (edges more pixely).
When rotation clockwise the color information is less sharp (edges smoothed).
Rotate the
[
Dyn.Set
]
digipot and select the desired dynamic range. Set 1 to Set 8.
1-4 color samples in axial direction shorter
4-8 color samples in axial direction larger
Remarks:
*
Changes are only possible in write mode.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
10-11
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...