Safety
2.13.8
Notes for Acoustic Output Tables for Track 3
Operating Conditions: means the adjustment of the scan parameters on the ultrasound console
MI:
is the Mechanical Index in an auto-scanning mode.
TISscan:
is the Soft Tissue Thermal Index in an auto-scanning mode.
TISnon-scan: is the Soft Tissue Thermal Index in a non-autoscanning mode.
TIB:
is the Bone Thermal Index.
TIC:
is the Cranial Thermal Index.
Aaprt:
is the area of the active aperture (square centimeters).
pr.3:
is the derated Peak Rarefactional Pressure (megapascals).
W0:
is the ultrasonic power, except for TISscan, in which case it is the ultrasonic power
passing through a one centimeter window (milliwatts).
W.3(z1):
is the derated ultrasonic power at axial distance z1.
ISPTA.3(z1): is the derated spatial-peak, temporal-average intensity at axial distance z1 (milliwatts
per square centimeters).
z1:
is the axial distance corresponding to the location of max[min(W.3(z).ITA.3(z) * 1
cm2)], where z³zbp (millimeters).
zbp:
is
1.69(Aaprt)
1/2.
For MI, zsp: is the axial distance at which pr.3 is measured; for TIB, zsp is the axial distance at
which TIB is a maximum (i.e. zsp = zb.3) (millimeters).
deq(z):
is the equivalent beam diameter as a function of axial distance z, and is equal to
[(4/p)(W0/ITA(z))]
1/2, where ITA(z) is the temporal-average intensity as a function of
z (millimeters).
fc:
is the center frequency (megahertz).
EBD:
are the entrance beam dimensions for the azimuthal and elevational planes
(millimeters).
1 - 12
PD:
is the pulse duration (microseconds).
PRF:
is the pulse repetition frequency (kilohertz).
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
2-17
Summary of Contents for Voluson 730Pro
Page 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Page 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Page 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Page 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Page 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Page 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Page 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...