Table 4-2: Discrete digital inputs at TB7
(continued)
TB7
Function
Pin 9
Digital input 5
Pin 10
Digital input return
Wire a ROC800 digital input (DI) module
To connect the ROC800 DI module to a field device:
Procedure
1. Expose the end of the wire to a maximum length of 0.25 in (6.4 mm).
NOTICE
We recommend twisted-pair cables for in/out (IO) signal wiring. The module's
terminal blocks accept wire sizes between 12 and 22 American wire gauge (AWG).
Allow some slack when making connections to prevent strain.
NOTICE
Failure to follow this notice may cause a short circuit and damage equipment. Allow
only a minimal amount of bare wire to prevent short circuits.
2. Insert the exposed end into the clamp beneath the termination screw.
3. Tighten the screw.
Reference Manual
Installation and start-up
2-3-9000-744
June 2022
Rosemount 700XA
77