4.4.16
Configure analytical train
Use the
Analytical Train Configuration
window for multiple analysis clocks to assign the
valves, discrete outputs (DO), detectors, electronic pressure controls (EPCs), and heaters
to each train and then assign each train to its respective analysis clock.
Figure 4-34: Analytical Train Configuration window
Procedure
1. Assign the usage of valves and DO to Analyzer# on
Hardware
→
Valves
,
Hardware
→
Detectors
,
Hardware
→
Discrete Outputs
,
Hardware
→
EPC
, and
Hardware
→
Heaters
screens.
2. Open the
Application
→
Analytical Train Configuration
screen.
You can use the
Filter Selections
drop-down list to filter by the type of hardware you
are interested in. The options are:
•
All
•
Detectors
•
Valves
•
Discrete Output
•
EPC
•
Heaters
By default, All is selected, and all types of hardware are displayed. To filter by a
certain type of hardware, select it from the list. Then only rows with the selected
hardware will be displayed.
3. Click
Discrete Output
and
Valves
. Assign the respective DOs, valves, and detectors
to each analytical train.
The valves are assigned with Usage as
Analyzer#
displayed on this screen. All
available detectors are also displayed on this screen. You cannot configure the same
Reference Manual
Installation and start-up
2-3-9000-744
June 2022
Rosemount 700XA
93
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......