WARNING
Burns
Some parts of the analyzer may be heated to 248 °F (120 °C).
To prevent burns, do not touch any of the hot parts. All parts of an analyzer are always hot
unless it has been switched off and allowed to cool down.
Before fitting, removing, or performing any maintenance on the analyzer, make sure
that it has been switched off and allowed to cool for at least two hours.
When handling the analyzer, always use suitable protective gloves.
These precautions are particularly important when working at heights.
If burned, seek medical treatment immediately.
Tools required
•
7 ⁄64-in. Allen wrench
•
Phillips screwdriver
•
Flathead screwdriver
Procedure
1. Disconnect all power to the unit.
Allow at least 10 minutes for the components to cool-down.
2. Remove the explosion-proof dome and the thermal hood.
Reference Manual
Operation and maintenance
2-3-9000-744
June 2022
Rosemount 700XA
137