10. Power up the GC and connect to it through MON2020.
11. In MON2020, go to
Chromatograph
→
View/Set Date_Time
. Set the date and time
for the GC.
Rosemount MON2020 Software for Gas Chromatographs Reference
12. In MON2020, go to
Tools
→
Cold Boot
. Cold boot the GC.
The GC reboots automatically and disconnects from MON2020.
13. Wait for the GC to reboot.
14. Reconnect to the GC using MON2020.
15. In MON2020, go to
File
→
Restore Configuration (to GC)
. Use the configuration
file you saved in
or use the last known good configuration.
16. Wait for the heaters to stabilize.
17. Go to
Control
→
Auto Sequence
to auto sequence the GC.
5.4.7
Repairing and maintaining the valves
Only minimal valve repair and maintenance is required (e.g., replacing the diaphragms).
Required tools for valve maintenance
The tools required for performing repair and general maintenance on the Rosemount XA
Series valve assemblies are:
•
Torque wrench, scaled in foot-pounds
•
½
-in. socket for 10-port valves
•
7/16-in. socket for 6-port valves
•
¼
-in. open-ended wrench
•
5/16-in. open-ended wrench
•
5/32-in. Allen wrench
Valve replacement parts
Replacement parts required for each Rosemount XA Series valves consist of the following
parts:
•
Diaphragm kit 6-port XA valve (PN 2-4-0710-248)
•
Diaphragm kit 10-port XA valve (PN 2-4-0710-171)
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
110
Emerson.com/Rosemount
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......