2.2.5
Archived data storage capabilities
Type of record
Number of records
Remarks
Analysis results
31,744
88 days with 4-minute cycle time
Final calibration results
370
1 year
Calibration results
100
N/A
Final validation results
370
1 year
Validation results
100
N/A
Analysis chromatograms
Approximately 22.5 days assuming running 4-
minute analysis and 1 analysis clock
Final calibration chromatograms
370
Final validation chromatograms
370
Protected chromatograms
100
User-selectable
Approximately 9 days, assuming 4-minute
cycle time
Daily averages
365
1 year
Weekly averages
58
1 year
Monthly averages
12
1 year
Variable averages
N/A
Every run (up to 250 variables)
N/A
Alarm logs
1000
N/A
Event logs
1000
N/A
(1)
Changed from 2.0.x release.
(2)
The gas chromatograph (GC) can store final calibration and validation chromatograms for a year, provided that no
more than one calibration/validation is run per day and the cycle time is less than 15 minutes. If the cycle time exceeds
15 minutes, the oldest final calibration/validation chromatograms are deleted to make room for newer ones.
(3)
You can have a total of 256 averages, including hourly, 24-hour, weekly, monthly, variable, and every run averages.
Equipment description and specifications
Reference Manual
June 2022
2-3-9000-744
42
Emerson.com/Rosemount
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......