2.2
Specifications
Type
Specification
Dimensions (without sampling
system)
Height x width x depth: 50 in (1,270 mm) x 40 in (1,016 mm) x 24 in (610 mm)
Weight (without sampling system)
Approximately 150 lb (68.0 kg)
Mounting
•
Wall mount (standard)
•
Free-standing (optional)
Power usage
125 to 250 W
Valve actuation
•
Sample gas: 90 psig (6 barg) maximum
•
Carrier gas: 90 psig (6 barg) maximum
•
Actuation gas: 110 psig (8 barg) maximum
Environment
Indoor/outdoor
Thermal conductivity detector (TCD): -4 °F (-20 °C) to 140 °F (60 °C)
Flame ionization detector (FID): 32 °F (0 °C) to 140 °F (60 °C)
Flame photometric detector (FPD): 32 °F (0 °C) to 122 °F (50 °C)
Hazardous area certifications
(hardware dependent)
USA and Canada
•
Class I, Zone 1, Ex/AEx db IIC, Gb T6/T4/T3
•
Class I, Division 1, Groups B, C, and D, IP66
EU ATEX and IECEx
•
Ex db IIC Gb T6/T4/T3
•
Ta = -4 °F (-20 °C) to 140 °F (60 °C)
•
SIRA 08ATEX 1328X
•
IECEx SIR 08.0093X
Consult factory for additional product certifications available.
Table 2-1: Approval temperature ratings
T6
Basic system; no alternative options included
T4
Liquid sample injection valve (LSIV) option included
T4
Heat trace option with a maximum 176 °F (80 °C) temperature switch set point
T3
Heat trace option with a maximum 230 °F (110 °C) temperature switch set point
Reference Manual
Equipment description and specifications
2-3-9000-744
June 2022
Rosemount 700XA
39
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......