2. Bring the two leads in through one of the two possible entries on the lower
compartment. Connect to the termination plug provided with the GC.
Figure 4-7: Wiring entries on the under side of the lower enclosure
Attribute
Wire color
+ (positive)
red
– (negative)
black
NOTICE
Do not disconnect the factory-installed ground wire.
The backplane board that connects to the 24 Vdc is protected from lead reversal by
the use of blocking diodes.
If the red (+) and black (-) leads are inadvertently reversed, no damage will occur;
however, the system will not have power.
3. Connect the DC power leads to the power disconnect switch that should be
properly fused.
The recommended fuse size is 8 amps.