6. On the new CPU board, set up switches as shown in the following image:
Figure 5-4: CPU switch settings
A. Turn SW7 ON (toward the dot).
B. Turn SW3 OFF (away from the dot).
C. Turn SW4 OFF (away from the dot).
D. Turn SW6 OFF (away from the dot).
Note
Rosemount 700XA GCs are tagged with CPU board part number
7A00555G02
.
7. Install the new CPU board in the card cage.
Ensure the board is seated firmly in place.
8. Place the clear plastic cover back over the boards.
9. Close the GC cover.
Reference Manual
Operation and maintenance
2-3-9000-744
June 2022
Rosemount 700XA
109
Summary of Contents for Rosemount 700XA