Figure 5-57: RS-232 connections
Terminal
Label
Definition
1
RX
Receive
2
TX
Transmit
3
RTS
Request to send
4
DTR
Data termininal ready
5
GND
Ground
Procedure
1. Start Rosemount MON2020 and connect to the GC.
2. Select
Tools
→
I/O Cards
.
3. Identify the appropriate card slot under the
Label
column and then select
Communications module - RS232
from the appropriate
Card Type
drop-down list.
4. Click
OK
.
5. Turn off the GC.
6. Install the RS-232 board into the appropriate I/O card slot in the GC’s card cage.
7. Close and secure the electronics enclosure door.
8. Apply power to start the GC.
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
172
Emerson.com/Rosemount