Wiring the discrete digital outputs
The discrete outputs are located on
TB3
, which is a 15-pin connector, and have five Form-
C relays on the backplane. All contact outputs have a rating of 1A at 30 Vdc.
Figure 4-21: TB3 on the backplane
lists the discrete digital output function for each pin on the
TB3
connector.
Table 4-4: Discrete digital outputs on TB3
TB3
Function
Pin 1
Normally closed (NC1)
DIG_OUT NC1
Pin 2
ARM1
DIG_OUT ARM1
Pin 3
Normally open (NO1)
DIG_OUT NO1
Pin 4
NC2
DIG_OUT NC2
Pin 5
ARM 2
DIG_OUT ARM2
Pin 6
NO2
DIG_OUT NO2
Pin 7
NC3
DIG_OUT NC3
Pin 8
ARM3
DIG_OUT ARM3
Reference Manual
Installation and start-up
2-3-9000-744
June 2022
Rosemount 700XA
79