4. If present, remove the clear cover from the card cage.
Figure 5-47: Card cage
5. Unplug all the cards in the card cage but do not remove them.
6. Unscrew the three switch panel connector posts. Remove the washers as well.
7. Lift the card cage with the boards and remove it from the lower enclosure.
8. Unscrew and remove the post closest to the power supply.
9. Unplug the connector at the top of the power supply on the left.
10. Unplug the low voltage cable connected along the lower edge of the backplane.
11. Disconnect the ground lead from the power supply at the chassis ground
immediately inside the lower enclosure opening.
12. Remove the nut just above the power supply. Twist the power supply free of the
attaching stud and lift it from its anchor plate. Remove the power supply carefully
to avoid damage due to wire interferences.
13. Using a Phillips screwdriver, remove the two screws attaching the anchor plate to
the inside of the enclosure.
14. Remove the anchor plate through the front opening of the enclosure.
Reference Manual
Operation and maintenance
2-3-9000-744
June 2022
Rosemount 700XA
161
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......