5. Connect the fiber cable to the µFPD PMT module and tighten the two cable clamp
Allen screws.
The shoulder of fiber cable shall be flush with clamp surface to ensure fiber cable is
fully inserted into PMT module.
Figure 5-36: Fiber cable connected to the µFPD PMT module
A. Fiber cable
B. Cable clamp
C. Cable clamp screws
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
150
Emerson.com/Rosemount