2. Grasp the edges of the FID board and align with the six socket tubes that extend
into the pins in the cap.
3. Replace the screw at the top of the FID assembly.
4. Tighten the tubing nut connector, securing the FID exhaust tube.
5. Screw in the tubing nut connectors located at the base of the FID.
6. Install the explosion-proof cover over the FID assembly and secure to the base with
eight bolts.
Replace igniter/thermocouple assembly
Figure 5-16: Igniter/thermocouple assembly front view
A. Bulkhead nut
B. Screws and washers
C. Fitting
D. Bulkhead nut
E. Igniter/thermocouple assembly
F. Igniter connector
Reference Manual
Operation and maintenance
2-3-9000-744
June 2022
Rosemount 700XA
127