4.4.2
Connect optional AC/DC power converter
WARNING
Failure to follow this warning may result in injury or death to personnel or cause damage
to equipment.
Check the gas chromatograph (GC) prior to wiring to determine if it is equipped for
optional AC power.
Procedure
1. Locate the plug-together termination block inside the electronics enclosure, atop
the power supply and adjacent to the card cage.
Figure 4-8: AC/DC termination block
WARNING
Failure to follow this warning may result in injury or death to personnel or cause
damage to equipment.
Do not connect the AC power leads without first ensuring that the AC power source
is switched
Off
.
NOTICE
Failure to observe this precaution may cause damage to equipment.
Do not apply electrical power to the GC until all interconnections and external signal
connections have been verified and proper grounds have been made.
AC wiring is usually color coded as:
Label
Wire color
Hot (H)
Brown or black
Neutral (N)
Blue or white
Installation and start-up
Reference Manual
June 2022
2-3-9000-744
58
Emerson.com/Rosemount
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......