Unless stated otherwise in the product documentation, ensure that the
pressure of the calibration and sample line is regulated at 10 to 30 psig (0.7
to 2.1 BarG). 15 psig (1 BarG) is recommended.
c) Validate calibration gas and retention times and run a manual calibration.
d) Go to
MON2020
→
Application
→
Component Data
and select the
associated stream. Check the
Component Data
table for calibration gas
validation information and retention times.
e) Go to
MON2020
→
Control
→
Calibration
and select the analysis stream to
run a manual calibration. Select the
Purge stream for 60 seconds
checkbox
and
Normal
calibration type radio button; then click
OK
.
NOTICE
Consult the
Rosemount MON2020 Software for Gas Chromatographs
for more information.
2. Select
Control
→
Auto Sequence...
to start auto sequencing of the line gas
stream(s).
NOTICE
Rosemount MON2020 Software for Gas Chromatographs Reference
The gas chromatograph (GC) begins the auto sequence analysis.
Reference Manual
Installation and start-up
2-3-9000-744
June 2022
Rosemount 700XA
99
Summary of Contents for Rosemount 700XA
Page 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Page 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Page 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Page 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Page 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Page 280: ......
Page 281: ......
Page 282: ......
Page 293: ......