6. Select
Hardware
→
Heaters...
.
The
Heaters
window displays.
Figure 4-35: Heaters window
7. Allow the GC system temperature to stabilize and the carrier gas lines to become
fully purged with carrier gas, which usually takes at least an hour.
The temperature values for the heaters should indicate that the GC is warming up.
The
Status
column displays
OK
.
8. Select
Control
→
Auto Sequence...
.
NOTICE
Rosemount MON2020 Software for Gas Chromatographs Reference
Note
with the local operator interface (LOI).
Important
Emerson recommends a continuous operation without sample gas for a period of
four to eight hours (or overnight), during which no changes should be made to the
settings described in
through
4.5.4
Purge calibration gas lines
Prerequisites
Purging calibration gas lines requires power and a personal computer (PC) connected to
the gas chromatograph (GC).
Reference Manual
Installation and start-up
2-3-9000-744
June 2022
Rosemount 700XA
97