Table 5-3: Configuring the serial connections for RS-485/RS-422 serial
communications
Jumpers
RS-485 (half duplex/2-wire)
RS-422 (full duplex/4-wire)
J3
Half
Full
J5
Half
Full
Termination IN
Termination OUT
J4
In
Out
J6
In
Out
TB1 wire terminals
RS485 (half duplex/2-wire)
RS422 (full duplex/4-wire)
A
RxTx+
Rx+
B
RxTx-
Rx-
Y
Normally closed
Tx+
Z
Normally closed
Tx-
5.4.13
Installing or replacing a F
OUNDATION
™
Fieldbus module
Figure 5-59: F
OUNDATION
Fieldbus module
The F
OUNDATION
Fieldbus module should be mounted adjoining the card cage. It is held in
place by the LOI post tips that attach to the LOI posts.
Mounting the F
OUNDATION
Fieldbus module requires the following items:
•
A F
OUNDATION
Fieldbus module
Operation and maintenance
Reference Manual
June 2022
2-3-9000-744
174
Emerson.com/Rosemount