Figure 5-14: FID
A. Termination board
B. Cap
C. Ground shield cover
D. Mounting bracket
E. Vent connector
F. Exhaust tube
Procedure
1. Disconnect all power to the gas chromatograph (GC).
Allow at least 10 minutes for the components to cool down.
2. Locate the FID switch, which is on the half-moon-shaped wire terminal board, and
flip it to the
OFF
position.
Reference Manual
Operation and maintenance
2-3-9000-744
June 2022
Rosemount 700XA
125